By David C. Eagleson, The Baker Company, Inc.
Class II biological safety cabinets (BSCs) are vital pieces of laboratory equipment in many life science applications requiring contamination control. In addition to providing a workstation with aseptic conditions for product protection, a BSC also helps protect laboratory personnel from exposure to aerosols of hazardous substances and prevents the release of such hazards into the environment. Field certification is the method by which proper BSC operation is verified over time, to be sure that product, personnel, and environmental protection are maintained.
Requirements for Class II biological safety cabinets in the United States are established by NSF International and published in NSF/ANSI Standard 49.1 Class I and Class III cabinets also exist and are generally used only for special applications. Class II cabinets are by far the most prevalent, and NSF 49 is specific to Class II BSCs. Therefore, this article will focus on Class II BSCs, and all subsequent uses of “BSC” are in respect to Class II BSCs.
NSF 49 includes definitions of the types and function, acceptable materials, design and construction requirements, and performance requirements for Class II BSCs. NSF International manages a program of type testing and product certification for Class II BSCs. Annex F of the standard includes specific requirements for field certification. To ensure that individuals performing certification are properly qualified, NSF International also administers a program for Biosafety Cabinet Field Certifier Accreditation.2
This article will review the function and designs of Class II BSCs and describe the field tests that should be performed during certification. The relationship between certification and commissioning will then be discussed, and the process and requirements for accreditation of field certifiers will be reviewed.
Types of Class II biological safety cabinets
All Class II biological safety cabinets offer product, personnel, and environmental protection from biological and other aerosolized contaminants. Product protection is offered by unidirectional (commonly called “laminar”) downflow air in the work chamber, generated by the cabinet blower pushing air through the supply HEPA filter. Personnel protection comes from the intake air pulled into the front access opening of the cabinet. Environmental protection is provided by HEPA filters in the exhaust air stream of the cabinet. The airflow pattern in a generic Class II BSC is shown in Fig. 1.
|
Different types of Class II BSCs are utilized when protection from chemical vapor hazards is also a concern, as HEPA filters do not capture gases or vapors.3,4 Type B1 and B2 cabinets must be directly connected to the building’s exhaust system for venting to the outdoors via a hard connection (see Fig. 2). Type A1 and A2 cabinets usually return their filtered exhaust air to the room but may optionally be connected to the building exhaust system with a canopy, formerly called a “thimble,” connection (see Fig. 3). Note that many existing installations of exhausted Type A cabinets may utilize a direct connection to the exhaust system. However, more recent revisions of the NSF 49 standard recommend use of the canopy exhaust, and in the future hard connections for Type A cabinets will most likely not be allowed at all. See Table 1 for a summary of the definitions of different cabinet types.
|
null
|
null
Field tests for biological safety cabinets
BSC field tests should be performed by the certifier upon installation and relocation of cabinets, after major maintenance is conducted, changing of HEPA filters, and at regular intervals thereafter. NSF 49 recommends a maximum interval of 12 months between certification, though many organizations re-certify more frequently. For example, pharmacies certify their BSCs every 6 months as required by USP Chapter <797> on sterile compounding.
|
In order to verify proper operation of a biological safety cabinet, the following tests are performed.1 These tests are related to the containment and product protection provided by the cabinet, and results must correlate to the values obtained by NSF for type testing of that particular make, model, and size of cabinet.
- Downflow velocity profile test
- Inflow velocity test
- Airflow smoke pattern tests
- HEPA filter leak test
- Cabinet integrity test (for Type A1 cabinets with positive-pressure contaminated plenums only)
- Alarm function verification
- Blower interlock (for Type B1 and B2 cabinets)
- Exhaust system performance (for any cabinet connected to the building exhaust)
While not part of Class II BSC field certification to NSF Standard 49, for certain installations additional validation tests may be required. For example, most cGMP compliant facilities will verify air cleanliness with a particle counter to the required ISO class per IEST protocols. Any such validation tests should be performed in conjunction with
field certification.
The following tests, related to worker comfort and safety, may optionally be performed. These tests verify functions of the cabinet not directly related to containment or product protection.
- Lighting intensity test
- Vibration test
- Noise level test
- Electrical tests (leakage, ground circuit resistance, and polarity)
Certification and commissioning
The goal of commissioning is to verify and document that the facility and its systems meet defined objectives and criteria for performance.5 Commissioning is performed prior to occupying new laboratory facilities to ensure that systems are operating as specified. Many organizations also repeat commissioning, or at least some portion thereof, at regular intervals to ensure continued proper operation of building systems. This “re-commissioning” is a good practice to guarantee that all building systems continue to function as required.
One of the most important systems considered in the commissioning process for laboratory facilities is the HVAC (heating, ventilation, and air conditioning) system. This is where commissioning and certification have a vital overlap and interdependency. Proper HVAC design and operation is crucial to the proper operation of BSCs. While the majority of BSC installations simply return cabinet exhaust to the room, they may still impact, or be impacted by, the function of the HVAC system.
From a heating and cooling standpoint, a typical 4-ft Class II, Type A2 cabinet may generate 2,000 to 3,000 BTUs/hr. Air conditioning and heating systems should be verified to maintain desired environmental conditions when BSCs are in operation, as well as when cabinets are turned off.
More crucial to the safety of laboratory workers and the integrity of aseptic work processes is the possible impact of air currents in the room on the performance of the BSC itself. Location of room air supplies and returns is critical, as cross-drafts may negatively affect the performance of BSCs.6 This dependency should be taken into account when designing room air ventilation and should be verified during the commissioning process.
Another level of complexity is introduced when BSCs are vented to the outdoors via the building exhaust system. With such an installation, the BSC itself becomes the first piece of the system ductwork and needs to be designed for and tested as such. In addition to the possible impact of cross-drafts, the commissioning process for BSCs should also verify that the required exhaust and room supply air to operate the BSC are available and that HVAC controls are interfaced with the BSC controls.
1. Supply air
Supply air requirements for BSCs are often overlooked. However, they can be just as critical as exhaust air requirements to proper cabinet function. Whatever volume of air is exhausted by the BSC must also be supplied to the room in order to avoid “starving” the cabinet of air. The supply air available to the BSC should be verified as well as the supply air to maintain desired room pressurization and air exchange rates. A BSC cannot be certified if a lack of supply air causes a low or inconsistent inflow or downflow velocity.
2. Exhaust airflow and static pressure
Biological safety cabinets are constant volumetric airflow devices, and the fan energy to exhaust the cabinet must be provided by the building exhaust system. In addition to the exhaust airflow, the static pressure requirements for hard connected Type B cabinets are relatively high (up to