Issue



Guidelines for selecting an optical particle counter (OPC)


12/01/2001







by Charles A. Pashby, Richard D. Barcus, and Michael T. Sloan

Click here to enlarge image

When it comes to choosing an OPC, consider sensitivity, counting efficiency, accuracy and reproductibility.

Selecting an optical particle counter (OPC) can appear deceptively simple. Typically, the specification focuses on sensitivity, flow rate, size range and coincidence loss. Secondary requirements are number of channels, the sample/hold periods and alarm limits. There are some serious hazards, however, when selecting an OPC on the basis of these parameters only.

For example, specifying sensitivity without considering the counting efficiency curve could mean good sensitivity (ability to sense small particles) but extremely poor resolution (ability to detect small differences in particle size). Poor resolution can cause large errors in the particle count. Relying exclusively on sensitivity or counting efficiency measurements based on ideal (transparent and spherical) test particles can result in wrong answers when counting particles in the real world.

These particles occur in a wide variety of shapes and refractive indices, causing large errors in particle sizing. Failure to recognize the difference between a "highly tuned" lab instrument (which can easily slip out of calibration with normal handling) and a "ruggedized" field instrument (which holds its calibration month after month) can be costly. Poor calibration stability causes sizing drift, non-repeatability, random spikes and, ultimately, loss of user confidence.

OPC theory
A light source (typically a plasma laser or laser diode) is collimated to illuminate a sample volume of aerosol flowing out of a nozzle. A photodetector, off-axis from the light beam, measures the amount of light scattered from single particles by refraction, reflection and diffraction. Both the size and the number of particles are measured simultaneously.

The size of the particle is deduced from the intensity of the scattered light. Although this article focuses on aerosol OPCs, the guidelines presented are equally applicable to liquid counters. In the case of a liquid OPC, the fluid is constrained to a channel inside a transparent cell (for example, quartz). Figure 1 shows the relationship between particle size and the amount of scattered light intensity when testing a typical OPC with monodisperse latex spheres.

In the Rayleigh region, where particles are smaller than the light wavelength, light is scattered equally in all directions (isotropically) from the particle. Its intensity varies as a function of the 6th power of particle size in this region (exp = 6).


Figure 1. A pictorial view of a typical OPC.
Click here to enlarge image

In the Mie region, where particles are nearly the same size as the light wavelength, the light pattern surrounding the particle becomes scalloped. The forward lobe (pointing in the same direction as the laser beam) becomes larger as the particle size increases. The exponential relationship changes inversely with particle size. In the Geometric region, where particles are much larger than a wavelength, classical optical theory takes over. Light scattering from a particle can be calculated from the physical effects of diffraction, reflection, refraction and absorption.

Important parameters are defined in terms of measurements based on suspended polystyrene latex (PSL) spheres. These parameters are examined for comparison to actual particles found in the cleanroom. For convenience, OPC measurements are based on the introduction of aerosol with PSL particles of highly monodispersed sizes over the range of approximately 0.1 to 3 microns. The one-sigma dispersion is typically ±1 percent. If we introduce PSL particles much larger than the specified lower detection limit, say 0.5 microns; the readings will be dispersed about a center value, as shown in Figure 2. Accuracy is the difference between the true value and the center value of the interval.

Accuracy: The "correctness" of

the size measurement; expressed as a percentage:

A = (DM - D1) x 100% / D1

where DM, is the measured diameter and D1, is the true diameter (see Figure 3).


Figure 2. If we introduce PSL particles much larger than the specified lower detection limit, say 0.5 microns, the readings will be dispersed about a center value.
Click here to enlarge image

OPC manufacturers generally specify sensitivity and counting efficiency on the basis of ideal test particles that are transparent and spherical. Most often, PSL spheres, with a refractive index of about 1.59, are used for testing. In the particle counting industry there is a tendency to emphasize PSL sphere sensitivity and skirt the issue of OPC sizing accuracy and sensitivity with particles found in the real world. Unfortunately, real-world particles come in a variety of shapes and refractive indices, leading to a significant degradation of sensitivity, resolution and accuracy. Sizing real-world particles is an inexact science.

Particle diameter
Resolution: The smallest detectable particle size difference. It is the ratio of the standard deviation (a) to the diameter (D) expressed as a percentage:

% Res = s x 100% / D.

Resolution is a function of the width of the bellshaped curve. It is also referred to as "coefficient of variation, relative precision and relative standard deviation."

Precision: The standard deviation (d) of repeated measurements of the same size monodispersed spheres where: D1 = the i-th measurement of particle diameter (arithmetic mean of N measures). N = total number of measurements.

Reproducibility (also called repeatability and calibration stability): The extent to which an OPC will give the same sizing and counting response to the same diameter PSL spheres over a long period of use.

Sensitivity: The smallest size particle an OPC can detect at a specified counting efficiency, for example, 0.3 micron at 50 percent counting efficiency.

Counting efficiency: The detected particle concentration divided by the true concentration (as measured by a hypothetically perfect instrument). This curve provides useful information regarding the sensitivity and resolution of the instrument.

False count rate: The counts per unit volume using perfectly filtered air at a specified flowrate.

Signal: The magnitude of the sensed scattered light produced only by the passing of a particle through the view volume. Size is deduced from the signal magnitude. Noise is the opposite of signal in that it is produced by anything but a particle in the view volume. A high signal-to-noise-ratio implies low false-count rate.

Counting efficiency, sensitivity and resolution
As an aid to arriving at a definition of counting efficiency, let's assume the presence of an ideal reference particle counter (see Figure 4). This counter can "see" every particle passing through the view volume to a diameter much lower than the lowest detection limit of the UPC under test. Typically, this instrument is a condensation nucleus counter (CNQ or an OPC) with a verified counting efficiency of 100 percent at the lower detection limit of the OPC under test. However, a CNC only counts particles above a given size corresponding to a preset threshold (for example, 0.01 micron); it cannot size particles. A reference CNC must be used with an electrostatic classifier to analyze particles by controlled deflection in an electrostatic field.

An aerosol carrying monodispersed PSL spheres is generated by the atomizer. The aerosol is mixed with filtered air in the mixing chamber. The OPC under test and the reference counter simultaneously sample the monodispersed spheres at the same concentration. As smaller and smaller monodispersed spheres are introduced, there is a point where the OPC under test fails to detect all the particles that the reference instrument is sensing. Further reduction in particle size results in the eventual loss of particle detection.

Counting efficiency is expressed as follows:

CE = Nm/N1 x 100%

where Nm is the measured concentration and N1 is the true concentration as measured by the reference instrument.

In all cases, the threshold of the counter's detector is set to sense those monodisperse particles which fall in the upper half of the bell curve (those particles to the right of the intersection point of the three curves shown Figure 5A). Smaller particles in the lower half of the curve (those to the left of the intersection point) are intentionally not sensed or counted.

Curve A represents the hypothetical case where the PSL spheres are ideal (exactly the same size with no dispersion) and the particles are sized perfectly by the OPC (also with no dispersion). Here, the 0, 50 and 100 percent efficiency points lie on the same vertical line.

In this case, the bell-shaped curve is simply a straight line. Unfortunately, particle counters do not exhibit such steep function efficiency curves.


Figure 3. Particle size vs. energy curve.
Click here to enlarge image

Curve B, exhibiting a relatively steep slope, is typical of a counter with good resolution, whereas curve C is representative of a counter with poor resolution. Size dispersion is much smaller in the case of the superior instrument (see curves in Figure 5B).

Let's examine the elements that determine the slope in practice. The largest contributor to poor resolution is the lack of uniformity of light intensity across the view volume. With any optical system, it's difficult to collimate a light beam down to a small area to achieve good sensitivity and, at the same time, maintain perfectly uniform intensity across this area. Any non-uniformity causes a discrepancy between sizing a particle passing right through the middle of the view volume and one that passes through one edge of the volume (being mistaken for a smaller particle). Variations in flow rate also contribute to wider dispersions resulting in degraded resolution. Other contributors, such as photodetector and amplifier stabilities are usually negligible in the typical particle counter.

If the counter thresholds corresponding to curves B and C had been set to anything but 50 percent, the particle counts between the two instruments would be in total disagreement. Only at 50 percent counting efficiency would two instruments with different resolutions count exactly half the monodispersed particles introduced.

If, for example, the thresholds were set for 90 percent counting efficiency at 0.12 micron as shown in Figure 5B, the curve C counter would outcount the curve B counter by a wide margin.

A quick and easy way to assess resolution is to compare the 50 and 100 percent points. In a typical MET One 0.1-micron counter, the difference is about 0.015 microns, corresponding to a resolution of about 5 percent. This represents good resolution in the OPC industry. A more effective approach is to obtain (or generate) the counting efficiency curve for the instrument in question and determine the size spread between the 10 and 90 percent points. This provides a solid basis for specifying and comparing OPCs.

Click here to enlarge image

In a typical cleanroom, the distribution of particles versus particle size follows a power law function as shown in the next curve. Given this distribution, theory predicts that a poor-resolution instrument will actually out-count one with good resolution at sizes below the threshold settings. Multichannel counters have a number of threshold settings, each set for 50 percent counting efficiency at the designated size. Note in the following set of curves the change in slope (and the apparent degradation of resolution) as particle size increases. This is due to the change in power-law exponent of the intensity of scattered light versus particle size.

In setting the threshold at the lower sensitivity level, the signal level must be above the noise level. If the threshold is set too close to the noise level, the false count rate (zero particles) will increase. To avoid this, always make sure that sensitivity and counting efficiency are accompanied by a minimum acceptable false-count specification.

Particles in the real world
Measurements with ideal PSL spheres provide us with a powerful tool for assessing the sensitivity, accuracy, resolution and false-count level of a counter. This calibration technique serves two purposes: 1) It gives comparative evaluations of a wide variety of counters and 2) provides a measure of how well a counter maintains its calibration (reproducibility).

Parameters such as accuracy, counting efficiency and resolution are very important in the process of PSL-based measurements. However, these parameters become meaningless when it comes to measuring real-world particles. Particles found in the cleanroom have a wide variety of shapes and refractive indices. This leads to a significant degradation of sensitivity, accuracy and resolution.

An article in the periodical, Microcontamination by Stuart Hoenig makes the point that "At 0. 1 microns, optical scattering...is strongly dependent on the particle's shape, color and electromagnetic characteristics," which reinforces the fact that significant error exists in sizing real world particles.[8]

To be able to "see" all of the particles in the view volume, the volume must be totally illuminated. Such a sensor is called "volumetric," one that is in-line with the flow and sees a sample of the aerosol is termed "in-situ." Some of the most sensitive counters available detect, for example, only 80 percent of the particles introduced and die curves are "normalized" to the 100 percent point when constructing the counting efficiency curves. All counting efficiency curves published by MET One reflect the true count. This fact must be considered when ordering a counter.

Some elements commonly found in contaminants, the by-products of processes like ionic etching and vacuum deposition, are:

  • Aluminum copper nickel
  • Boron fluoride phosphorus
  • Calcium gold potassium
  • Carbon iron sodium
  • Chlorine lithium sulfur
  • Chromium manganese tin

These elements exhibit a wide range of refractive indices. Some are highly reflective while others absorb most of the incident light energy.

In the case of carbon particles—which are highly absorptive—the deviation in respect to the in = 2.4 curve increases rapidly for particles above 0.2 µ, with the error approaching 250% above I µ.

For practical purposes, OPCs are calibrated with ideal particles having a refractive index between 1.5 and 1.6. The size measured by the OPC is then an "equivalent PSL diameter" or an "equivalent DOP diameter," depending on the calibrating aerosol used.

The magnitude of error in sizing real-world particles with an OPC would appear to be a discouragement in attempting to set up an effective cleanroom quality-assurance program. However, the outlook is not as bleak as might be expected. It turns out that the OPC can function as a surprisingly effective tool in the cleanroom if used in a protocol that has been evolved by cleanroom professionals over the years. The OPC has two basic functions in the cleanroom. The first is to certify the cleanroom to meet standards established by FED-STD-209D. The second function is to support a quality maintenance program in the cleanroom.

Air-quality maintenance
Monitoring air quality with the particle counter to support an air-quality maintenance program is more involved than cleanroom certification. The goal is to eliminate "killer defect" particles that can destroy product yield. Particles whose size is about one-tenth (or larger) of the minimum line width on a semiconductor wafer fall into this category.

Users must be more aware of the magnitude of particle sizing errors in the real world, due to variations in shape and refractive index. Instead of trying to size particles precisely, establish particle concentration reference levels and correlate these levels with product yield. The exact sensitivity (whether it was 0.1 or 0.15 µm), as measured earlier with PSL spheres, becomes insignificant. What is important is the ability of the counter to hold its calibration over the long term.

During a particle-shedding event, particles are generated in a wide distribution of sizes. If there are 0.1-micron particles in the sample, you can be sure there will also be 0.2-micron particles present. Considering this factor, the second decimal place in measuring micron sensitivity with PSL spheres should not be overemphasized. Calibration stability is more important.

Click here to enlarge image

Step 1 in a typical cleanroom monitoring procedure establishes a reliable zero count (false count level) using a quality filter on the counter's aerosol inlet port. Consistently low counts verify that you have attained an acceptable level. Step 2 establishes a particle concentration baseline for each station being monitored in the cleanroom. Even though the exact size of particles counted is unknown because of differences in shape and refractive indices, a reference level can be created. Familiarize yourself with the levels associated with the various processes and determine empirically what level is acceptable and what levels begin to reduce the yield. Step 3 periodically recalibrates with PSL spheres until you acquire confidence in the counter's ability to hold its calibration in the working environment.

Instead of an absolute particle sizer and counter, the OPC can be used more effectively as an early warning trend indicator or burst detector. This will allow you to shut down a process if the concentration level exceeds a preset threshold. Thus the OPC can function more as a process tool than an environmental tool in your cleanroom quality-assurance program.

Stability considerations
Stability in the working environment must be as important as sensitivity. It is extremely important that the counter maintain its calibration over the long term, otherwise, particle concentration baselines become meaningless.

Typically, the wavelength of light used in OPCs ranges from about 0.63 to 0.83 micron. As particles become smaller than a wavelength, the amount of light they scatter into the detector collection optics drops off rapidly. At 0.3 micron the detected energy drops off as a function of about the 4th power of particle size; at 0. 1 micron the detected light energy drops off as a function of about the 5th to 6th power of particle size.

Upgrading the sensitivity of a counter from 0.2 to 0.1 micron requires about a 17-fold increase in light power focused into the view volume. To achieve a sensitivity approaching 0.1 micron requires a well-designed laser/optical system with a narrow optical bandwidth (sometimes referred to as "high Q") in order to develop high light intensity in the view volume. Making the optical bandwidth too narrow (Q too high) in order to achieve high sensitivity can actually lead to calibration instability (in the presence of mild shock or vibration) with the attendant loss of sensitivity.

Each manufacturer is faced with making a trade-off between high sensitivity and reproducibility in the working environment. MET One, for example, puts the emphasis on stability and shock-resistant reproducibility in a tough environment at the expense of a slight sacrifice in sensitivity.

If a manufacturer decides to make the trade-off favoring sensitivity at the expense of calibration stability, the result is a sensitive, "highly tuned" laboratory instrument that can easily slip out of calibration with normal use. Simply moving such an instrument from one bench to another can degrade its detection-limit performance.

There is a tendency to tout OPC performance solely in terms of PSL sphere micron sensitivity at 50 percent counting efficiency (for example, 0.10 micron @ 50 percent counting efficiency). In the world of particle counting in a working environment, sensitivity alone is not a meaningful number. It does not give you enough information to determine if you are dealing with a highly tuned laboratory instrument or a durable and reliable workhorse.

Considering the degradation in particle counter sensitivity when dealing with real-world particles, MET One does not put undue emphasis on the second decimal place of the particle size sensitivity specification. Our position is clear: Acounter that maintains a long-term sensitivity of, for example, 0.12 microns (50 percent C.E.) in a rough environment is far superior to an instrument that begins with a sensitivity of 0.10 microns (50 percent C.E.) and slips out of calibration the first time it is moved to a new location.

Charles A. Pashby, is a applications engineer, Richard D. Barcus, is a applications engineer, Michael T. Sloan, is a technical writer for Grants Pass, OR-based Pacific Scientific Instruments. For more information on particle counting basics, contact Ken Szewc, air products manager, at [email protected] or Andy Young, liquids product manager, at [email protected].

References

  1. Martens, A.E., "Errors in Measurements and Counting of Particles Using Light Scattering," APCA Journal, Vol. 18, No. 10, October 1968.
  2. Quenzel, H., "Influence of Refractive Index on the Accuracy of Size Determination of Aerosol Particles with Light-Scattering Aerosol Counters," Applied Optics, Vol. No. 8, No.1, Ian, 1969.
  3. Willeke, K, Liu, B.Y.H., "Single Particle Optical Counter Principle and Application," University of Minnesota, Particle Technology Lab Publication No. 264, May, 1975.
  4. Burkman, D.C. et al, "Understanding and Specifying the Sources and Effects of Surface Contamination in Semiconductor Processing," Microcontamination, p. 57, November, 1988.
  5. Ranade, M.B.,"Properties of Airborne Particles," Prof. Dev. Course, University of Maryland, Mach 1989.
  6. Whitby K.T, Vomela, R.A., "Response of Single Particle Optical Counters to Nonideal Particles," Environmental Science and Technology, Vol. 1, Oct, 1967.
  7. Hoenig, S.A., "Where Are We and Where Do We Go Next?," Microcontamination, p.22, April 1989.
  8. Cooke, D.D., and M. Kerker, "Response Calculations For Light Scattering Aerosol Particles," Applied Optics Vol. 14 pp.734-739, March 1975.
  9. Procedural Standards for Certified Testing of Cleanrooms," National Environmental Balancing Bureau, October, 1988.
  10. "Federal Standard, Cleanroom and Workstation Requirements, Controlled Environment," (FEDSTD-2091), June 15, 1988.
  11. Hodkinson, J.R., "The Optical Measurement of Aerosols," Aerosol Science, ch. Xi Academic Press, 1966.